Basic Info.
Product Description


 
   | Main technical parameters ( water-cooled, temperature +25ºC, no load) 
 | |
| 1.Temperature range | -70ºC ~ +180ºC(Different temperature range per customer requirements) 
 | 
| 2.Temperature deviation | ±2.0ºC | 
| 3.Temperature fluctuation | <=1.0ºC | 
| 4.Temperature uniformity | <=±2.0ºC | 
| 5.Heating up rate&Cooling down rate 
 | from 25ºC to +150ºC whole process average>=3ºC/min Load 50 kg aluminum alloy 
 from +25ºC to -40ºCwhole process average >=2ºC/min Load 50kg aluminum alloy | 
| The temperature performance test is based on the relevant provisions of the IEC 60068 -3 
 standard; the sensor is placed at the unit air outlet. | |
| Chamber structure 
 | |
| The test chamber adopts the overall structure, the upper part is the working chamber, the rear side is 
 the refrigeration unit, the lower right space (>=700mm) is vacant to install the hydraulic part device, the 
 right is the electric control cabinet, the control panel is located on the right front side; the width direction 
 opens the door , the right side is the air duct system; the air supply motor, the centrifugal fan blade, the 
 heater, and the evaporator. | |

 
         